PicoScope 9200 USB / LAN Sampling Oscilloscopes

PicoScope 9200 Oscilloscope Specifications

Channels (vertical)
Channels 2 (simultaneous acquisition)
Bandwidth DC to 12 GHz
Pulse response rise time 29.2 ps
RMS noise, maximum <2.5 mV
Scale factors (sensitivity) 2 mV/div to 500 mV/div
Nominal input impedance (50 ±1) Ω
Input connectors SMA (F)
Timebase (horizontal)
Timebases 10 ps/div to 2 ms/div (main, intensified, two delayed, or dual delayed)
Delta time interval accuracy ±0.4% of delta time interval ±15 ps ±100 ppm of delay setting
Time interval resolution 200 fs minimum
Trigger
Trigger sources External direct trigger, external prescaled trigger, internal clock trigger, clock recovery trigger (9211 only)
Direct trigger bandwidth and sensitivity 100 mV p-p DC to 100 MHz, increasing linearly from 100 mV p-p at 100 MHz to 400 mV p-p at 1 GHz
Prescaled trigger bandwidth and sensitivity 200 mV p-p to 2 V p-p from 1 GHz to 8 GHz, 300 mV p-p to 1 V p-p to 10 GHz
Trigger RMS jitter, maximum 3.5 ps 20 ppm of delay setting
Data Acquisition
ADC resolution 16 bits
Digitising rate DC to 100 kHz maximum
Acquisition modes Sample (normal), average, envelope
Data record length 32 to 4096 points maximum per channel in x2 sequence
Display
Display resolution Variable
Display style Dots, vectors, variable persistence, infinite persistence, variable grey scaling, infinite grey scaling, variable colour grading, infinite colour grading
Measurement and Analysis
Marker Vertical bars, horizontal bars (measure volts) or waveform markers (x and )
Automatic measurements Up to 40 automatic pulse measurements
Histogram Vertical or horizontal
Mathematics Up to four math waveforms can be defined and displayed
FFT Up to two fast Fourier transforms can be run simultaneously with the built in filters (Rectangular, Nicolson, Hanning, Flattop, Blackman-Harris and Kaiser-Bessel)
Eye diagram Automatically characterises NRZ and RZ eye patterns. Measurements are based on statistical analysis of the waveform.
Mask test Acquired signals are tested for fit outside areas defined by up to eight polygons. Standard or user-defined masks can be selected.
Clock Recovery (PicoScope 9211 only)
Sensitivity 50 mV p-p typ from 12.3 Mb/s to 2.7 Gb/s continuous rate
Recover clock RMS trigger jitter, maximum 1.0% of unit interval
Maximum safe trigger input voltage ±2 V (DC peak AC)
Trigger input connector SMA (F)
PC Requirements
Processor Pentium-class processor or equivalent
Memory 256 MB
Disk space PicoScope 9000 software requires aproximately 30 MB
Operating system 32-bit edition of Windows XP (SP2) or Vista
Ports USB compliant port minimum.
USB 2.0 compliant port recommended.
Environmental
Operating environment
Temperature range

Humidity

5 °C to 35 °C for normal operation
15 °C to 30 °C for quoted accuracy
Up to 85% RH, non-condensing, at 25 °C
Storage environment
Temperature range
Humidity

-20 °C to 50 °C
Up to 95% RH, non-condensing
Physical Dimensions
Dimensions 170 x 255 x 40 mm   (6.7 x 10.0 x 1.6 in)
Weight 1.0 kg   (2.2 lb)
Language support
Documentation
User’s guide
Quick start guide

English
English
General
Supplied software/hardware PicoScope 9000 Series software
2 x SMA M-F adapters/connector savers
USB 2.0 cable
LAN cable (PicoScope 9211 only)
AC adaptor
Tough carry case
PC connection USB 2.0 (USB 1.1 compatible) - cable supplied
LAN connection 10/100 Mb/s - cable supplied (PicoScope 9211 only)
Power supply
AC adaptor
6 V DC ±5%. 1.9 A max (2.3 A for PicoScope 9211)
External adaptor supplied (suitable for USA, UK, Europe and Australasia)
Compliance FCC (EMC), CE (EMC and LVD)
Total Satisfaction Guarantee In the event that this product does not fully meet your requirements you can return it for an exchange or refund. To claim, the product must be returned in good condition within 14 days.
Warranty 2 years
Ordering Information
Variant PicoScope 9201 PicoScope 9211
Cat number PP463 PP473
Lead time Due to increased demand there is currently a 12 week lead time. Orders will be fulfilled in the order they are received. No payments are taken until goods are ready to be despatched.
 
 
 
USA Office

Tel:1.877.902.2979-1.425.223.4311
Fax:1.877.329.4324
Address: 1480 Gulf Road, Suite 837,
PO Box 1364
Point Roberts, WA 98281

Western Canada - Vancouver BC

Tel:1.800.663.6001 or 1.604.925.6150
Fax:1.604.925.6170
Address: 2454 Haywood Ave
West Vancouver, BC V7V 1Y1

Eastern Canada - Markham, Ontario

Tel:1.800.465.0164 or 1.905.513.7027
Fax:1.877.329.4324